Transferability studies for the AACC reference method and the IFCC method for measurement of alkaline phosphatase activity.

N W Tietz, A D Rinker, C A Burtis, P H Duncan, Kerry M. Ervin, Lilian M. Ewen, M. Mathieu, Claude Petitclerc, Daniel W. Grisley

PubMed · 1984 · 15 citations · 0 references

Concepts

Abstract

We present the results of measurements of alkaline phosphatase activity from interlaboratory transferability studies conducted by 12 laboratories in five countries. The variability, as demonstrated by within-day precision (CV less than or equal to 2.6%), between-day precision (CV less than or equal to 3.6%), and between-lab precision (CV less than or equal to 6.3%) establishes the transferability of this method for alkaline phosphatase. Some common errors encountered in enzyme measurements that affect the accuracy and precision of the measurements are listed.