Radioimmunoassay of Aldosterone in Plasma and Urine: Validation of a Novel Separation Technique and a Rapid Urine Assay

T.P. Jowett, Jeremy D. Slater, R. D. Piyasena, Roger Ekins

Clinical Science · 1973 · 31 citations · 13 references

Concepts

Abstract

1. A rapid, robust, accurate and sensitive radioimmunoassay procedure is described for the measurement of plasma and urinary aldosterone. 2. A new, simple method is validated for the separation of free from bound moieties which has general applicability and which involves the measurement of free steroid by direct extraction into toluene-based liquid scintillator. 3. A new method, of validated specificity, is described for the measurement of the acid-labile metabolite of aldosterone in urine which avoids the necessity for preliminary chromatographic purification.

References