Health ProfessionsPediatricsPhysiologyBioinstrumentationMedical PhysicsPhysical ExaminationCardiovascular MonitoringMedical InstrumentationVital SignsVital Signs AssessmentAssessment Of Oxygen SaturationOxygen Saturation MeasurementOxygen Saturation MonitoringPulse OximetersPulse OximetryPulse Oximetry Interpretation
The accuracy of the Ohmeda Biox 3700 and the Nellcor N100E was assessed in 25 cyanosed children. The readings obtained from the two pulse oximeters were compared with arterial blood measurements using a Radiometer OSM-2 co-oximeter. Both pulse oximeters differed significantly from the co-oximeter measurements and in these patients the error of both machines exceeded the manufacturers' claims. However, the machines appeared to reflect changes in saturation accurately in the same patient.
8
Mark Yelderman, William New · Anesthesiology · 1983 · 487 citations · Full text
Evaluation of Biox IIA ear oximeter.
P.M. Tweeddale, N J Douglas · Thorax · 1985 · 33 citations · Full text