Paediatrics & Child Health · 2011 · 16 citations · 10 references
MedicinePediatricsRadiologyHepatologyNeonatal HealthNeonatologyNeonatal Intensive CarePediatric Liver DiseaseMedical Records ReviewNeonatal Intensive Care Unit AdmissionUmbilical CatheterizationUmbilical Venous Access
In newborns, the presence of liver fluid collection is a rare event. The reported cases are isolated or described over long periods. Within four months, five neonates were diagnosed with liver fluid collection from safety occurrence reports. Clinical, laboratory and radiological data were extracted from medical records. The definite diagnosis was made by ultrasound. Four of the patients were preterm, male and had very low birth weights. The osmolality of the infused solution was within the acceptable range. Investigations revealed that the use of a new brand of umbilical vein catheter introduced in the neonatal intensive care unit, one month before the first case, was associated with this cluster. Low positioning of the umbilical vein catheter tip appeared to be a second contributory factor. Neonatal practitioners may benefit from the present report when facing the occurrence of similar lesions.
10
E. Simeunovic, Marion Arnold, Daniel Sidler et al. · Pediatric Surgery International · 2008 · 68 citations
Mal-Positioned Umbilical Venous Catheter Causing Liver Abscess in a Preterm Infant
Hugh Simon Lam, A. M. Li, Chiu‐Wing Winnie Chu et al. · Neonatology · 2005 · 51 citations
Neonatal total parenteral nutrition ascites from liver erosion by umbilical vein catheters
Brian D. Coley, John Seguin, Leandro Cordero et al. · Pediatric Radiology · 1998 · 49 citations
Medicine, Nutrition, Pediatrics +15