Improved manual and automated colorimetric determination of serum glucose, with use of hexokinase and glucose-6-phosphate dehydrogenase.

Holly J. Coburn, James J. Carroll

PubMed · 1973 · 56 citations · 0 references

Concepts

Abstract

A previously reported assay for serum glucose, in which hexokinase and glucose-6-phosphate dehydrogenase were used, was improved to replace the light-sensitive phenazine methosulfate with diaphorase, providing a one-reagent system. The assay system was automated for use with standard AutoAnalyzer equipment and evaluated at 40 and 60 samples per hour with a 2:1 sample-to-wash ratio. Sensitivity, precision, accuracy, and sample interaction were quite satisfactory at both sample rates. The assay benefits from the inherent advantages of the reaction mechanism: sensitivity, specificity, NADP regeneration, and freedom from dialysis