Life SciencesImmunologyInfectious DiseasesClinical ImmunologyDiagnostic MicrobiologyVirologyDiagnostic ImmunologyFalse PositivesHiv DiagnosisHiv TestingImmunodeficiency DisordersViral DiagnosticsFalse Negative Hiv ResultsFalse Positive Hiv ResultsHiv Antibody DetectionHiv Antigen DetectionHiv Serologic TestingHuman Immunodeficiency Virus TestingLaboratory Diagnosis Of HivWestern Blot Analysis
A case of a false-positive result of human immunodeficiency virus (HIV) confirmatory immunoblot-based assay is described. Repeatedly borderline reactive anti-HIV screening enzyme immunoassay result obtained in a local hospital resulted in directing the sample to the Slovenian HIV/AIDS Reference Laboratory. In the Reference Laboratory, both anti-HIV screening assays and confirmatory Western blot were negative, while a confirmatory test INNO-LIA HIV I/II Score (Innogenetics, Ghent, Belgium) was anti-HIV-1 positive due to sgp120 and gp41 reactivity. The results of serological testing of the second sample obtained three weeks later were completely identical, while in the third sample obtained 5 months later, seroreversion was observed. Due to a negative dynamics in anti-HIV serological profile and repeatedly negative results of the molecular tests for HIV-1 and HIV-2, HIV infection was excluded and the results of test INNO-LIA HIV I/II Score were finally interpreted as false positive.