Screening for retinopathy of prematurity employing the retcam 120: sensitivity and specificity.

118

Citations

10

References

2001

Year

Abstract

The sensitivity of the method was low. The ROP that was missed was peripheral stage 1 or stage 2 disease in peripheral zone 2 or zone 3. This was largely due to the technical limitations of the speculum-camera interface preventing a better view of the periphery. The ROP cases that were missed by the photographic examination regressed spontaneously on follow-up. No disease more posterior to peripheral zone 2 was overlooked. These results detail the accuracy of the method employing the technique of photoscreening as a potential substitute for detailed ophthalmological examination. At present there are clear technical limitations to such a substitution. The study is part of an ongoing project to determine the feasibility of employing neonatal nurses trained to take digitized images of the premature infant's retina and telemeter the results to be read by an experienced ophthalmologist remote from the site.