Publication | Closed Access
Partial-volume correction in PET: validation of an iterative postreconstruction method with phantom and patient data.
159
Citations
25
References
2007
Year
Iterative postreconstruction PVE correction generated more accurate uptake measurements in subcentimeter tumors for both phantoms and patients than the uncorrected values. The method eliminates the requirement for segmenting anatomic data and estimating tumor metabolic size or tumor background level. This technique applies a PVE correction to the mean voxel value within a VOI, yielding a more accurate estimate of uptake than the maximum voxel value.