Publication | Closed Access
A generic high-dose rate<sup>192</sup>Ir brachytherapy source for evaluation of model-based dose calculations beyond the TG-43 formalism
83
Citations
42
References
2015
Year
A hypothetical, generic HDR (192)Ir source was designed and implemented in two commercially available TPSs employing different MBDCAs. Reference dose distributions for this source were benchmarked and used for the evaluation of MBDCA calculations employing a virtual, cubic water phantom in the form of a CT DICOM image series. The implementation of a generic source of identical design in all TPSs using MBDCAs is an important step toward supporting univocal commissioning procedures and direct comparisons between TPSs.