Measuring the level of agreement between directly measured blood pressure and pressure readings obtained with a veterinary‐specific oscillometric unit in anesthetized dogs
28
Citations
17
References
2013
Year
There was poor agreement between the direct and indirect measured blood pressure measurement systems. Therefore, use of the oscillometric blood pressure unit evaluated in this study for monitoring patients under anesthesia cannot be recommended at this time.