Surface and Interface Analysis · 2014 · 44 citations · 11 references
Natural SciencesChemistryMaterials ScienceAnalytical ChemistryPolymer ChemistrySurface ChemistryChemical Composition AnalysisInstrumental AnalysisMass SpectrometryMass Spectrometry Data AnalysisPolymer AnalysisSurface AnalysisMass Spectrometric IdentificationMultivariate Curve FittingPolymer Mass SpectrometryPolymer Surface AnalysisSecondary Ion Mass SpectrometrySurface ContaminationSurface Contamination AnalysisTime-of-flight Secondary Ion Mass Spectrometry
Time‐of‐flight secondary ion mass spectrometry (ToF‐SIMS) is a powerful tool for determining surface information of complex systems such as polymers and biological materials. However, the interpretation of ToF‐SIMS raw data is often difficult. Multivariate analysis has become effective methods for the interpretation of ToF‐SIMS data. Some of multivariate analysis methods such as principal component analysis and multivariate curve resolution are useful for simplifying ToF‐SIMS data consisting of many components to that explained by a smaller number of components. In this study, the ToF‐SIMS data of four layers of three polymers was analyzed using these analysis methods. The information acquired by using each method was compared in terms of the spatial distribution of the polymers and identification. Moreover, in order to investigate the influence of surface contamination, the ToF‐SIMS data before and after Ar cluster ion beam sputtering was compared. As a result, materials in the sample of multiple components, including unknown contaminants, were distinguished. Copyright © 2014 John Wiley & Sons, Ltd.
11