Pacing and Clinical Electrophysiology · 1986 · 39 citations · 7 references
MedicineCardiologyPhysiologyElectrophysiologyExercise PhysiologyExercise TestingBiopotential MeasurementCardiac PacingCardiac PhysiologyCardiopulmonary Exercise TestingCardiovascular MonitoringMedical InstrumentationHeart Rate MonitoringImpedance MeasurementPacemaker PhysiologyPhysiological Parameter EstimationRate-responsive PacingRespiratory ImpedanceRespiratory Impedance MeasurementRight Ventricular Physiology
A rate responsive cardiac pacemaker will benefit the patient only if the input parameter is a reliable indicator of exercise. Part of the reliability is dependent upon selection of the best physiological parameter and part is dependent upon the measurement technique. Twenty-six patients have been studied in order to ascertain the best measurement configuration for an intravascular impedance respiratory monitor. Four electrode configurations have been examined at rest and during exercise. The results have been compared using measures of sensitivity and specificity. The ideal configuration requires that a current field be established between the right ventricular blood and the pacemaker case. And, that voltage sensing be done between an electrode in the SVC and the pacemaker case.