An Experimental Evaluation of Device/Arterial Wall Compliance Mismatch for Four Stent-Graft Devices and a Multi-layer Flow Modulator Device for the Treatment of Abdominal Aortic Aneurysms

74

Citations

29

References

2015

Year

Abstract

Commercially available endovascular devices lower the aortic wall compliance after implantation. The MFM was found to be the most compliant in the suprarenal region, while the Fortron device was the most compliant in the infrarenal region. Choosing the most compliant devices for treating AAAs produces positive gains in the aortic elastic recoil, thus minimising the device related complications.