Homocysteine and other thiols determined in plasma by HPLC and thiol-specific postcolumn derivatization

Anders Andersson, Anders Isaksson, Lars Brattström, Björn Hultberg

Clinical Chemistry · 1993 · 248 citations · 0 references

Concepts

Abstract

We describe a versatile high-performance liquid-chromatographic method for determining homocysteine and other plasma sulfhydryls. Using three different procedures for preparation of plasma, we determined total, free (non-protein-bound), and reduced forms of homocysteine, cysteine, glutathione, cysteinylglycine, and gamma-glutamylcysteine in human plasma. Sample preparation involves disulfide reduction with dithiothreitol and protein precipitation with sulfosalicylic acid. The assay utilizes isocratic reversed-phase ion-pair liquid chromatography at pH 2.4, postcolumn derivatization with 4,4'-dithiodipyridine, and colorimetric detection at 324 nm. The intra-assay precision (CV) of the method for total homocysteine is 1.5%; the interassay precision over 2.5 months is 2.5%. The detection limit for homocysteine is < 50 nmol/L plasma.