Concepedia

Publication | Open Access

2D measurements of cup orientation are less reliable than 3D measurements

13

Citations

29

References

2015

Year

Abstract

Errors in version in 2D were due to the difficulty in delineating the cup rim, which was obscured by a large-diameter metal head of the same radio-opacity. This can be overcome with 3D analysis. The present study demonstrates that measurements using EBRA have poor agreement and are less reliable than those with 3D-CT when measuring cup version and inclination in MoM hips.