PubMed · 1986 · 19 citations · 8 references
MedicineNuclear MedicineRadiologyCardiac Nuclear MedicineDiagnostic ImagingDiagnostic RadiologyCardiac ImagingCardiac Nuclear ImagingCardiovascular ImagingCoronary Artery ImagingMyocardial Perfusion ImagingPositron Emission TomographyCardiac Myocardial Perfusion ImagingCardiac Pet ImagingMyocardial Perfusion ScintigraphyNitrogen-13 Ammonia ImagingNitrogen-13 Ammonia Perfusion ImagingNitrogen-13 Ammonia QuantificationPositron Emission Tomography DetectionPositron Emission Tomography Myocardial Imaging
The effect of low sensitivity areas or gaps between adjacent slices of the multislice positron emission tomography on detection of myocardial perfusion abnormality with 13NH3 was evaluated segmentally in 36 patients with coronary artery disease at rest or during exercise. The detectability of the defects in RCA or LAD region was 80% in single-position scans in stress studies. The false-negative defects were located mainly in the inferior wall, apicoinferior wall, or high anterior wall. When the patients were moved half the slice interval to perform the interpolating scan, and the two sets of images were interlaced with each other, the detectability increased to 88%. The interpolating scan also allowed reconstruction of long-axis and short-axis tomograms in high quality, which further improved the detectability of perfusion defects (100% for RCA or LAD and 75% for LCX lesion) and helped in understanding the anatomic relationships to the coronary artery territories.
8
Anil V. Shah, Heinrich R. Schelbert, Markus Schwaiger et al. · Journal of the American College of Cardiology · 1985 · 112 citations
Medicine, Nuclear Medicine, Veterinary Diagnostic Imaging +14