Publication | Closed Access
Noise, Artifact, and Oversensing Related Inappropriate ICD Shock Evaluation: ALTITUDE NOISE Study
61
Citations
19
References
2012
Year
External noise and lead/connector noise were the primary causes, while T-wave oversensing was the least common cause of NAO resulting in ICD shock. Noise/artifact decreased immediately after a shock in nearly half of episodes. The specific ICD lead type did not impact the likelihood of NAO.