Publication | Closed Access
Microphacoemulsification with WhiteStar
87
Citations
13
References
2002
Year
Using the Sovereign phacoemulsification unit with the WhiteStar system during 2-stab-incision bimanual microphacoemulsification with a bare 19-gauge aspiration needle in human cadaver eyes, a wound burn could not be produced at the highest energy settings unless all flow into the eye and all aspiration were occluded. These settings were well beyond clinically applicable conditions and provide evidence that microphacoemulsification can be performed safely with the WhiteStar system.