The Journal of Urology · 1993 · 80 citations · 11 references
MedicineSurgeryImplant SurgeryProstheticsReconstructive SurgeryComplications Of Implant ProsthesesImplant ComplicationsImplant SurvivalPenile SurgeryProsthetic DevicesLife Table ParametersLong-term Implant SurvivalPenile ImplantationPenile Prosthesis ImplantationPenile Prosthesis SurgeryProsthesis Complications
A total of 214 patients underwent implantation of the AMS 700 penile prosthesis with the controlled expansion cylinders. Followup ranged from 9 to 86 months, with a mean of 56 months. Life table analysis to calculate the probability of the cylinder and the device as a whole surviving 6 years revealed 97% for the cylinder and 90% for the device as a whole. No aneurysmal dilatation of the cylinders occurred. The incidence of cylinder leak was 0.7%. The infection rate in patients undergoing primary implantation of the device was 0.5% compared to 6.6% in patients undergoing revision surgery. The controlled expansion cylinders have significantly reduced the incidence of mechanical failures and, thus, the need for reoperation. As a consequence of the decreased need for reoperation the incidence of device infection has also decreased.
11
Estimating survival functions from the life table
Edmund A. Gehan · Journal of Chronic Diseases · 1969 · 259 citations
Experience with inflatable penile prosthesis
Irving J. Fishman, F. Brantley Scott, J. Keith Light · Urology · 1984 · 51 citations
Medicine, Surgery, Prosthetics +8
Clinical experience with scott inflatable penile prosthesis in 150 patients
Daniel C. Merrill · Urology · 1983 · 37 citations
Medicine, Surgery, Prosthetics +12