Concepedia

Publication | Closed Access

Objective and subjective evaluations of the Nurotron Venus cochlear implant system via animal experiments and clinical trials

14

Citations

24

References

2015

Year

Abstract

In the animal and cadaveric head experiments, magnet dislocation could not be prevented in the 1.5 T MRI without removal of the internal magnet. The EABR was clearly elicited in cat experiment. In the clinical trial, the ESRT was strongly correlated with C level (p < 0.001). The human clinical trial involving 57 post-lingually deafened native Mandarin-speaking patients was performed. Residual hearing protection in the implanted ear at each audiometric frequency was observed in 27.5-46.3% patients post-operatively. A pitch ranking test revealed that place pitches were generally ordered from apical to basal electrodes. The recognitions of the perceptions of 301 disyllabic words, environment sounds, disyllabic words, and numerals were significantly better than the pre-operative performance and reached plateaus.

References

YearCitations