Publication | Closed Access
The Porous Coated Anatomic Total Hip Replacement
115
Citations
17
References
2001
Year
There have been persistent problems with the Porous Coated Anatomic hip system, including thigh pain and an increasing prevalence of osteolysis with time. Revision because of aseptic loosening was related more to the thickness of the polyethylene liner than to the size of the femoral head. Femoral heads with a 32-mm diameter did not increase the risk for revision provided that an adequate thickness of polyethylene had been used.
| Year | Citations | |
|---|---|---|
Page 1