Journal of Forensic Sciences · 2010 · 82 citations · 25 references
Chemical Warfare AgentForensic ChemistryIon Mobility SpectrometryChemistryGas ChromatographyExplosive CompoundsBioanalysisLinear Dynamic RangeAnalytical ChemistryLiquid ChromatographyClinical ChemistryInstrumentationBiophysicsChromatographyDrug AnalysisBiochemistryChromatographic AnalysisPharmacologyIon MobilityNatural SciencesExplosive ChemistryMass SpectrometryMedicineGc⁄ Ims
Ion mobility spectrometry (IMS) is a valued field detection technology because of its speed and high sensitivity, but IMS cannot easily resolve analytes of interest within mixtures. Coupling gas chromatography (GC) to IMS adds a separation capability to resolve complex matrices. A GC-IONSCAN® operated in IMS and GC⁄ IMS modes was evaluated with combinations of five explosives and four interferents. In 100 explosive/interferent combinations, IMS yielded 21 false positives while GC⁄ IMS substantially reduced the occurrence of false positives to one. In addition, the results indicate that through redesign or modification of the preconcentrator there would be significant advantages to using GC⁄ IMS, such as enhancement of the linear dynamic range (LDR) in some situations. By balancing sensitivity with LDR, GC⁄ IMS could prove to be a very advantageous tool when addressing real world complex mixture situations.
25
Herbert H. Hill, William F. Siems, Robert H. St. Louis et al. · Analytical Chemistry · 1990 · 244 citations
Engineering, Biological Mass Spectrometry, Altmetric Attention Score +17