Concepedia

Publication | Closed Access

Decrease in power output of new light-emitting diode (LED) curing devices with increasing distance to filling surface.

49

Citations

0

References

2002

Year

Abstract

The blue LED curing devices Elipar FreeLight and GC e-Light showed a significantly higher percentage decrease in power output at a distance of 10 mm from the light guide tip to the radiometer compared to the QTH curing devices. Therefore, although blue LED curing devices might have the same curing potential compared to a QTH device when placed in direct contact to a resin composite, blue LED curing devices may not provide a sufficient cure when placed at a clinically relevant distance of 10 mm to the resin composite surface.