A Suggested Standard Method for the Investigation of Electrophoresis

Laurence S. Moyer

Journal of Bacteriology · 1936 · 74 citations · 12 references

DOIFull text

Open access

Concepts

Abstract

Although the number of papers on electrophoresis is rapidly increasing, the errors in technique to be found in many of them would seem to make a discussion of the necessary safeguards in using such methods desirable. Numerous techniques are avail-able in the literature but the one described in this paper has been used with such uniform results by several investigators, over a period of ten years, that it seems advisable to advance it as a possible standard procedure and to summarize in one place the method in some detail. Many of the developments involved have not been previously described. THE ELECTROPHORESIS CELL Choice of cell. For quantitative studies, microelectrophoresis cells are to be preferred to the U-tube type in regard to simplicity of theory and ease of operation (Abramson, 1932, 1934a). Both flat and cylindrical microelectrophoresis instruments are available but, of the two, the flat cell is better suited for routine measure-ments. The theory of the cylindrical cell is slightly simpler but to use it properly an optical bench is required. In addition, errors in leveling are hard to eliminate because the "stationary" level is a horizontal cylinder in form. This means that particles in the same microscope field are at different levels unless the focus is directed exactly on the high point of the cylinder. The flat cell is much easier to handle. The author knows of three types of flat cells in general use which can be made to yield good results in proper hands. These

References

12