ORIGINAL ARTICLE: Under‐filled blood collection tubes containing K<sub>2</sub>EDTA as anticoagulant are acceptable for automated complete blood counts, white blood cell differential, and reticulocyte count
26
Citations
6
References
2009
Year
Blood CellOriginal ArticleReticulocyte CountLaboratory HematologyHematologyClinical ChemistryLaboratory MedicineRadiologyHealth SciencesBecton DickinsonTransfusion MedicineBlood CollectionCurrent Laboratory StandardsBlood DonationCoagulopathyMedicineAnticoagulantBlood TransfusionEmergency Medicine
Current laboratory standards from Clinical Laboratory Standards Institute (CLSI) and manufacturer's (Becton Dickinson) data indicate that under-filling K(2)EDTA blood collection tubes can result in erroneous hematology values. To accommodate under-filled tubes and reduce collection volumes while optimizing our automation, we explored the acceptable limit of under-filled tubes for hematology values. We collected 8.0 ml of blood from 30 normal adult volunteers. Each donation was aliquoted in the following volumes: 4.0, 2.0, 1.0, 0.5 ml x 2. These samples were analyzed within 1 h of blood collection on Sysmex XE-2100 (Sysmex America Inc., Mundelein, IL, USA) for complete blood count, reticulocyte, and white blood cell differentials. Results of the under-filled tubes were compared to those of the standard volume. The Deming regression analysis show excellent correlation for all parameters between each under-filled blood collection volume compared to a standard 4 ml volume. The Bland and Altman analysis shows good agreement between both 1.0 and 2.0 ml compared to a 4.0 ml volume. The 0.5 ml compared to a 4.0 ml volume, however, shows increased variation on many parameters. In addition all three collection volumes show negative bias compared to the standard volume for platelet count, but the difference is considered insignificant with a percent difference of 5.5%, 3.2%, and 1.5% for 0.5, 1.0, and 2.0 ml collection volume respectively. Finally for 0.5 ml collection volume we noticed a low level of false positive flagging rate for white blood cell. Acceptable complete blood count values of under-filled powdered K(2)EDTA tubes can be obtained with as little as 1.0 ml of blood.
| Year | Citations | |
|---|---|---|
Page 1
Page 1