EngineeringElectrical EngineeringAnalog ElectronicsInstrumentationMetrologyAnalog InstrumentationCurrent Measurement InstrumentsElectrical MeasurementElectrical MetrologyNoise In Analog CircuitsNoise MeasurementAccuracyCurrent MeasurementCurrent Measurement AccuracyCurrent NoiseCurrentsDirect Current MetrologyMeasurement System CapabilityOn-site Calibration
An interlaboratory comparison of small-current generation and measurement capability is presented with the ultrastable low-noise current amplifier (ULCA) acting as travelling standard. Various measurements at direct currents between 0.16 nA and 13 nA were performed to verify the degree of agreement between the three national metrology institutes involved in the study. Consistency well within one part per million (ppm) was found. Due to harsh environmental conditions during shipment, the ULCA's transfer accuracy had been limited to about +/-0.4 ppm. Supplemental measurements performed at PTB indicate that further improvements in accuracy are possible. Relative uncertainties of 0.1 ppm are achieved by applying on-site calibration of the ULCA with a suitable cryogenic current comparator.