PubMed · 2004 · 17 citations · 0 references
PharmacyPharmaceutical AnalysisPharmacologyAnalytical ToxicologyBioavailabilityChromatographic AnalysisChromatographic Analysis Of PharmaceuticalsChromatographyLiquid ChromatographyPreclinical Drug EvaluationToxicological Screening Of PharmaceuticalsAssayAssay Method ValidationBioanalytical ChromatographyPurity CharacterizationPurity TestingReversed-phase ChromatographyReversed-phase Gradient ElutionTablet Impurity Analysis
A reversed-phase gradient HPLC method was developed for the evaluation of pioglitazone hydrochloride (PG-HCl) in tablets. Limit of detection for PG-HCl was found to be 42 ng/ml. Analyses were performed on a C18 column (Symmetry C18, 5 microm, 250 x 4.6 mm), mobile phase was a mixture of ammonium formate buffer adjusted with formic acid to pH 4.1 and acetonitrile. Shortened purity method was used as the assay method. Methods were validated.