Brightness alters Heidelberg retinal flowmeter measurements in an in vitro model.

39

Citations

5

References

1999

Year

Abstract

The HRF may report velocities inaccurately because of varying brightness in the fundus. In the present experiment, a darker background produced an overreporting of velocities. An offset, possibly introduced by a noise correction routine, apparently contributed to the inaccuracies of the HRF measurements. Such offsets vary with local and global brightness. Therefore, HRF measurements may be error prone when comparing eyes. When used to track perfusion in a single eye over time, meaningful comparison may be possible if meticulous care is taken to align vessels and intensity controls to achieve a similar level of noise correction between measurements.