EducationKinesiologyMovement AnalysisPhysical ActivitySleep MedicineSleep AssessmentSleep BehaviorSleep HealthSleep ManagementSleep PhysiologySleep Signal AnalysisSensitivitySleep EfficiencySleep Quality AssessmentSleep TimeSleep-wake Rhythm MeasurementSleepiness MeasurementSubjective Sleep MeasurementWearable Activity Monitoring
STUDY OBJECTIVES: To compare 2 commercially available actigraphs. DESIGN: Subjects wore the Actiwatch L and the Basic Mini-Motionlogger for 2 nights. SETTING: Naturalistic. PARTICIPANTS: 20 healthcare workers aged 24 to 64 years. MEASUREMENTS: Measures included sleep latency, total sleep time, wake after sleep onset, and sleep efficiency. Actiwatch L data were analyzed at 3 sensitivity settings. RESULTS: The Mini-Motionlogger reported more total sleep time, less wake after sleep onset, and greater sleep efficiency than the Actiwatch L set at high sensitivity. It also reported more wake after sleep onset than the Actiwatch L set at low sensitivity. CONCLUSIONS: Both devices achieved similar overall performance when the Actiwatch L was set at medium sensitivity.
5
Automatic Sleep/Wake Identification From Wrist Activity
Roger J. Cole, Daniel F. Kripke, William Gruen et al. · SLEEP · 1992 · 1.9K citations · Full text
An Activity-Based Sleep Monitor System for Ambulatory Use
John Webster, Daniel F. Kripke, Sam Messin et al. · SLEEP · 1982 · 281 citations · Full text
Wrist-Actigraphic Estimation of Sleep Time
Daniel J. Mullaney, Daniel F. Kripke, Sam Messin · SLEEP · 1980 · 242 citations · Full text