Publication | Open Access
Comparison of uncalibrated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output measurements
121
Citations
22
References
2006
Year
Arterial waveform analysis with an uncalibrated algorithm COWave underestimated COPAC to a clinically relevant extent. The wide range of LOAs requires further evaluation. Better results might be achieved with an improved new algorithm. In contrast to this, we observed a better correlation of thermodilution COTranspulm and thermodilution COPAC measurements prior, during, and after CABG surgery.