MedicineBiomedical EngineeringBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyAttenuation CorrectionData NormalizationData TransformationDose NormalizationImage PreprocessingPositron Emission TomographyImage NormalizationImaging Result NormalizationIterative ImprovementPartial Volume CorrectionPet Attenuation CorrectionPet Image ReconstructionPet QuantificationSystem Normalization
In this work an iterative ML technique is developed to normalise acquired PET data. We proposed a model for component-based correction featuring geometric, crystal efficiency and block timing factors. The algorithm is tested against the conventional fan-sum method and with a non-ML iterative technique on both simulated and acquired data. The results show that the iterative methods are superior to the conventional fan-sum technique. Furthermore the new method provides an improved normalisation over the previously published iterative technique when low statistics acquisitions are used.