Validity of Acoustic Reflectometry in Detecting Middle Ear Effusion

Daniel M. Schwartz, Richard H. Schwartz

PEDIATRICS · 1987 · 37 citations · 0 references

Concepts

Abstract

Pneumootoscopy, tympanometry, and acoustic reflectometry were performed in 256 middle-class children seen in a surburban pediatric office. The results demonstrated that relectometry, when validated by otoscopic findings, detected middle ear effusion with 88% sensitivity and 83% specificity when a cut-off of 5 linear units was used. Corresponding values for tympanometry were 87% and 77.5%. These results are in keeping with those of earlier studies on acoustic reflectometry and demonstrate the usefulness of this simple technique in detecting chronic and acute otitis media with effusion.