Publication | Open Access
Impact of positional changes in neural monitoring endotracheal tube on amplitude and latency of electromyographic response in monitored thyroid surgery: Results from the Porcine Experiment
54
Citations
10
References
2015
Year
Amplitude changes without latency changes may be due to changes in tube position alone during surgery, but could still reflect a neurophysiologic event; amplitude changes during neuropraxic injury merit additional investigation. Thus, the combined event (concordant amplitude decrease and latency increase) serves as an appropriate adverse EMG event correlating with impending neural injury. © 2015 Wiley Periodicals, Inc. Head Neck 38: E1004-E1008, 2016.
| Year | Citations | |
|---|---|---|
Page 1