IEEE Transactions on Nuclear Science · 1998 · 34 citations · 6 references
MedicineNuclear MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyDetector TechnologyFiber OpticsOptical FibersPositron Emission TomographyScintillation DetectorsScintillator MaterialsCoincidence DetectionDetectorsFibersMultilayer StacksPhoton DetectionPositron Emission Tomography DetectorsRadiation Energy ResolutionTime Resolution
The authors have constructed and tested prototype PET detector modules incorporating wavelength-shifting fiber ribbons alternating with thin CsI(Na) scintillator plates in multilayer stacks. Modules with cross-sectional area of 11 cm/spl times/11 cm and with 9 mm of total crystal thickness have been tested. A reconstructed spatial resolution of 3 mm FWHM has been obtained for a line source, with an energy resolution of <20% and a time resolution of 25 ns FWHM. Energy measurements and coincidence timing are provided by 4 photomultipliers in an Auger array on one 11 cm/spl times/11 cm module face. Multiplexed fiber readout has been implemented with multianode photomultipliers coupled to several fiber ribbons. The methods and resolution achieved are readily extended to larger systems in a straightforward and cost-effective manner. Results of tests with several phantoms are presented.
6
Feasibility study for positron emission mammography
Christopher J. Thompson, K. Murthy, I. Weinberg et al. · Medical Physics · 1994 · 170 citations