PubMed · 1995 · 13 citations · 0 references
MedicineCardiologySurgeryCardiothoracic SurgeryInterventional CardiologyPediatric CardiologyBalloon AngioplastyCongenital Heart DiseaseCoronary AngioplastyPercutaneous Coronary InterventionThoracic SurgeryBalloon Catheter DilationBalloon Pulmonary AngioplastyCoronary Balloon AngioplastyPulmonary Artery Balloon DilationPulmonary AtresiaRadiofrequency TreatmentVentricular Septum
Six days old neonate with pulmonary atresia and intact ventricular septum is reported in which a communication was established between the right ventricle and the pulmonary artery. The membraneous atresia was perforated with a special radiofrequency catheter 2F connected to a Hat 200S radiofrequency generator (Osypka). The energy necessary to perforation was 15W. After perforation of the atresia, balloon dilation was performed (balloon diameter: 2, 3, 4mm of coronary angioplasty and 5, 8mm Penta Medi-tech). The valve was completely opened giving the continuity between right ventricle and pulmonary artery.