Intravascular optical coherence tomography: optimisation of image acquisition and quantitative assessment of stent strut apposition.

Jun Tanigawa, Peter Barlis, Carlo Di Mario

PubMed · 2007 · 161 citations · 0 references

Concepts

TL;DR

Stent expansion, apposition, and symmetry are key criteria for optimal deployment, yet intravascular ultrasound underestimates malapposition (~7%) while optical coherence tomography offers superior resolution (10–15 µm) and fewer artifacts. This study aims to describe methods and techniques that enable optimal assessment of stent strut apposition with the current OCT system. Using a motorised pullback OCT system that requires transient proximal flow occlusion, the authors acquire contiguous high‑resolution images to overcome IVUS limitations and address the learning curve for image quality. Immediate malapposition after drug‑eluting stent implantation was not linked to increased adverse events, indicating that early concerns about impaired drug delivery and stent failure are unfounded.

Abstract

Stent expansion, apposition and symmetry were the three criteria of intravascular ultrasound (IVUS) guided optimal stent deployment in the bare metal stent era1, with at least the criterion of stent expansion maintaining its clinical relevance in the drug-eluting stent (DES) era2. Two prospective studies showed that stent malapposition immediately following DES implantation was not associated with increased adverse clinical events3,4. Thus, initial concerns that immediate stent malapposition would affect drug delivery to the vessel wall and lead to DES failure appeared to be unfounded. Coronary IVUS, however, formerly the gold standard for assessing stent strut apposition, is imprecise in the detection of stent malapposition observed in around 7% of lesions3,4, a figure which is probably grossly underestimated. This is due to the limited axial and lateral resolution of the ultrasound waves (100-150 microm) and the constant presence of artefacts around stent struts (side-lobes, shadowing). Optical coherence tomography (OCT) uses infrared light with the advantage of greater resolution (10-15 microm) and less strut induced artefacts compared with IVUS5,6. Unlike the first OCT prototypes used by Jang et al5,6, new commercial systems (LightLab Imaging Inc., Westford, MA, USA) can acquire contiguous images with a motorised pullback system as used with IVUS7. Still, OCT has some disadvantages including a more complex process of image acquisition that requires transient proximal flow occlusion with a balloon. Since there is definite learning curve to achieve optimal image quality, this article will focus on methods and techniques to attain optimal assessment of stent strut apposition, using the current OCT system.