Publication | Closed Access
Human <i>in vivo</i> phosphate metabolite imaging with <sup>31</sup>P NMR
143
Citations
31
References
1988
Year
Phosphorus (31P) spectroscopic images showing the distribution of high-energy phosphate metabolites in the human brain have been obtained at 1.5 T in scan times of 8.5 to 34 min at 27 and 64 cm3 spatial resolution using pulsed phase-encoding gradient magnetic fields and three-dimensional Fourier transform (3DFT) techniques. Data were acquired as free induction decays with a quadrature volume NMR detection coil of a truncated geometry designed to optimize the signal-to-noise ratio on the coil axis on the assumption that the sample noise represents the dominant noise source, and self-shielded magnetic field gradient coils to minimize eddy-current effects. The images permit comparison of metabolic data acquired simultaneously from different locations in the brain, as well as metabolite quantification by inclusion of a vial containing a standard of known 31P concentration in the image array. Values for the NMR visible adenosine triphosphate in three individuals were about 3 mM of tissue. The ratio of NMR detectable phosphocreatine to ATP in brain was 1.15 +/- 0.17 SD in these experiments. Potential sources of random and systematic error in these and other 31P measurements are identified.