Evaluation of an Automated Colony Counter

William A. Goss, R. N. Michaud, Michael B. McGrath

Applied Microbiology · 1974 · 23 citations · 1 references

DOIFull text

Open access

Concepts

Abstract

An automated colony counter was found to readily detect surface and subsurface bacterial colonies of 0.3-mm size or greater with a high degree of precision. On a logarithmic scale, counting efficiency consistently ranged from 89 to 95% of corresponding manual count determinations for plates containing up to 1,000 colonies. In routine application, however, automated plate counts up to approximately 400 colonies were selected as a more practical range for operation. The automated counter was easily interfaced with an automated data acquisition system.

References

1