Publication | Closed Access
Effects of antibiotic-impregnated external ventricular drains on bacterial culture results: an in vitro analysis
52
Citations
19
References
2009
Year
The findings in the current study indicate that the risk of a false-negative culture result may be increased when a CSF sample is drawn through an AI catheter. In the management of a known shunt infection, a false-negative result from an EVD culture specimen may lead to an inappropriately short duration of antibiotic therapy. These data have significant clinical implications, particularly given the widespread use of AI drains and the current high rates of shunt reinfection after EVD use worldwide.