Publication | Closed Access
<sup>236</sup> U AMS measurement at CIRCE
21
Citations
9
References
2010
Year
Measurement TheoryEngineeringNuclear PhysicsMeasurementMeasurement Standards (Educational Assessment)EducationIsotopic Abundance SensitivityCalibrationApplied MeasurementAnalytical ChemistryInstrumentationIsotope AnalysisPrecision MeasurementAccelerator Mass Spectrometry16-Strip Silicon DetectorIsotope GeochemistryMass SpectrometryIsotopic RatioMeasurement Standards (Electrical Engineering)Measurement System
In order to measure the isotopic ratio of actinides, the upgrade of the accelerator mass spectrometry system at the Center for Isotopic Research on Cultural and Environmental heritage at the Second University of Naples, Italy, was performed. The beam emittance of 238U and the isotopic abundance sensitivity of 236U were measured on the present beam line. Utilizing a 16-strip silicon detector, the sensitivity of 236U/238U≈1×10−11 was obtained.