Relative 99mTc-MAG3 renal uptake: reproducibility and accuracy.
50
Citations
0
References
1999
Year
The methods that combined the best reproducibility and accuracy for estimating 99mTc-MAG3 left-to-right uptake ratio were the integral method, with subrenal or perirenal background correction, and the Patlak-Rutland plot. The use of the integral method without background correction introduced a systematic bias, whereas the slope method resulted in high variability. Therefore these methods cannot be recommended.