Concepedia

Publication | Closed Access

Systematic evaluation of the root cause of non‐linearity in liquid chromatography/tandem mass spectrometry bioanalytical assays and strategy to predict and extend the linear standard curve range

49

Citations

25

References

2012

Year

Abstract

In this work, the root cause of the non-linear behavior of the standard curve when using a SIL-IS was investigated and identified. Based on the findings, an improved multiple SRM channels approach was proposed and successfully applied to obtain a linear dynamic range of five orders of magnitude for one test compound. This approach may work particularly well for LC/MS/MS bioanalytical assay of dried blood spot (DBS) samples, for which a direct dilution is cumbersome.

References

YearCitations

Page 1