2015 · 10 citations · 8 references
EngineeringCell BiologyHematologyAnalytical ChemistryInstrumentationMetrologyBioanalytical Method ValidationMeasurement UncertaintyCell CountingCell Density MeasurementDensity MeasurementsGravimetric MeasurementParticle Count MeasurementRandom CoincidencesSample Volume MeasurementSystematic Uncertainties
A method is described to determine cell concentrations for erythrocytes and leukocytes with typical uncertainties of 0.8% and 2%, respectively. The unique method is designed in such a way that results are traceable to SI derived units. The sample suspensions are prepared under gravimetric control. The sample volume measured in cell counting is determined by weighing, including separate density measurements. Random coincidences are a significant challenge in particle counting leading to underestimated cell concentrations. In the method described here this challenge is met by recording the integrated dead time in particle counting and applying dilution series. This method is now routinely applied to provide reference measurement values for external quality assurance of German hematology laboratories for the complete blood count.
8
Photo-Electric Technique for the Counting of Microscopical Cells
Andrew Moldavan · Science · 1934 · 168 citations
Medicine, Biology, Cell Biology +6
Carol Schnizlein‐Bick, John Spritzler, Cynthia L. Wilkening et al. · Clinical and Diagnostic Laboratory Immunology · 2000 · 118 citations · Full text
Life Sciences, Immunology, Pathology +12
Keith A. Reimann, Maurice R.G. O’Gorman, John Spritzler et al. · Clinical and Diagnostic Laboratory Immunology · 2000 · 100 citations · Full text