Anaesthesia · 2010 · 17 citations · 4 references
MedicineOtolaryngologyRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyAirway AssessmentAirway DisordersAirway ImagingAirway ObstructionLaryngeal ImagingMagnetic Resonance ImagingAirway Device DesignFerromagnetic ResonanceLaryngeal Mask Airway DevicesMagnetic ResonanceMagnetic Resonance PhysicsPhantom DevicesSupraglottic ImagingTracheal Imaging
We investigated the artefacts created during magnetic resonance imaging by five different laryngeal mask airways: the Classic (cLMA); the LMA ProSeal; the LMA Unique; the Ambu Disposable Laryngeal Mask; the LMA Supreme; and one other supraglottic airway device, the i-gel supraglottic airway. The devices were placed on top of and inside a phantom simulator to resemble the position in vivo. The artefacts with the cLMA, Unique and Supreme were similar and related to ferromagnetic material in the pilot balloon valve. Artefacts were more prominent with the ProSeal. There were no artefacts with the Ambu Disposable Laryngeal Mask or the i-gel.
4
Laryngeal Mask Anesthesia: Principles and Practice
Tohru Asai · British Journal of Anaesthesia · 2005 · 102 citations · Full text
Study of the adjustment of the Ambu laryngeal mask under magnetic resonance imaging
E. Monclús, Antonio Garcés, Belén De José María et al. · Pediatric Anesthesia · 2007 · 11 citations · Full text
Medicine, Anesthesiology, Radiology +17