Journal of Neurotrauma · 2002 · 63 citations · 19 references
MedicineNeurological SurgeryNeurosurgeryNeurocritical CareSensorsTrauma NeurosurgeryCerebral Blood FlowIntracranial HypertensionIntracranial PressureIntracranial Pressure ManagementIntracranial Pressure MonitoringNeurologic MonitoringCerebrospinal Fluid Pressure MeasurementComplication Rate AnalysisContinuous Intracranial Pressure MonitoringIncreased Intracranial PressureIntracranial Pressure Measurement AccuracyIntracranial Pressure Monitoring DevicesInvasive Brain MonitoringNoninvasive Intracranial Pressure Monitoring
To assess the safety and accuracy of the Camino intraparenchymal sensor, we prospectively evaluated hemorrhagic complications, zero-drift, infection, and system malfunction in 163 patients monitored after a severe head injury. Mean duration of intracranial pressure (ICP) monitoring was 5 +/- 2.2 days (range: 12 h to 11 days). Of the 141 patients with a control CT scan, four showed a 1-2-cc collection of blood at the catheter's end. When removed, the sensors underread the true ICP value (negative zero-drift) in 80 of the 126 sensors evaluated (63.5%). Fourteen sensors showed no zero-drift, and 32 sensors overread the true ICP value (positive zero-drift) (median: -1 mm Hg; interquartile range: -4 to +1 mm Hg). No significant relationship was found between zero-drift, the surgeon who implanted the sensor, intracranial hypertension, or duration of ICP monitoring. No clinical infections could be attributed to the devices. Sixteen patients (9.8%) required more than one ICP sensor due to malfunctioning of the system. In conclusion, continuous ICP monitoring using the Camino intraparenchymal sensor has a low complication rate. However, this sensor may underread the real ICP values in a high number of patients. The lack of correlation between duration of ICP monitoring and zero-drift suggests that, contrary to the recommendations of other reports, the intraparenchymatous Camino sensor can provide reliable readings after the fifth day of use.
19
Intracranial Pressure Monitoring and Liver Transplantation for Fulminant Hepatic Failure
Steven D. Lidofsky, Nathan M. Bass, Marie Csete Prager et al. · Hepatology · 1992 · 265 citations · Full text
The Incidence and Significance of Hemostatic Abnormalities in Patients with Head Injuries
John D. Olson, Howard H. Kaufman, Joel L. Moake et al. · Neurosurgery · 1989 · 210 citations
Medicine, Hematology, Neurosurgery +14