Publication | Closed Access
Correlation of Anterior Chamber Angle and Ciliary Sulcus Diameters With White-to-White Corneal Diameter in High Myopes Using Artemis VHF Digital Ultrasound
111
Citations
21
References
2009
Year
Regression modeling found weak correlations among all combinations of white-to-white, angle diameter, and sulcus diameter. Given the relative accuracy of direct measurement of angle and sulcus diameter compared to the potential accuracy of these regression equations, it appears that direct measurement would increase the safety of anterior and posterior chamber phakic intraocular lens sizing.