Journal of Biomedical Materials Research Part B Applied Biomaterials · 2015 · 11 citations · 28 references
Natural SciencesBiomedical EngineeringMaterials ScienceBiomaterialsNanomaterialsSurgical OncologyAblation TherapyBiomaterials For Tissue EngineeringLiver Tissue EngineeringNanomedicineElectrosurgical ElectrodesElectrosurgical UnitsLiver Tumor AblationThermal AblationThermal Ablation EnhancementThermal EffectsThermal Injury ResponseThermal Properties
The objective of this study was to use surface functionalization to evaluate the antiadhesion property and thermal injury effects on the liver when using a novel electrosurgical unit with nanostructured-doped diamond-like carbon (DLC-Cu) thin films for tumor ablations. The physical and chemical properties of DLC-Cu thin films were characterized by contact angle goniometer, scanning electron microscope, and transmission electron microscope. Three-dimensional (3D) hepatic models were reconstructed using magnetic resonance imaging to simulate a clinical electrosurgical operation. The results indicated a significant increase of the contact angle on the nanostructured DLC-Cu thin films, and the antiadhesion properties were also observed in an animal model. Furthermore, the surgical temperature in the DLC-Cu electrosurgical unit was found to be significantly lower than the untreated unit when analyzed using 3D models and thermal images. In addition, DLC-Cu electrodes caused a relatively small injury area and lateral thermal effect. The results indicated that the nanostructured DLC-Cu thin film coating reduced excessive thermal injury and tissue adherence effect in the liver.