IEEE Symposium Conference Record Nuclear Science 2004. · 2005 · 11 citations · 2 references
EngineeringNuclear MedicinePhysicsParticle PhysicsRadiation DetectionRadiation PhysicsGamma-ray DetectionGamma-ray DetectorsParticle DetectorsPositron Emission TomographyRadiation DetectorsScintillation DetectorsCherenkov DetectionCherenkov DetectorsPet Detector TechnologyPhotoelectric EffectPositron Emission Tomography DetectionPositron Emission Tomography DetectorsScintillationTime Resolution
Silica aerogel with the refractive index of around 1.2 has been developed. It is argued that a Cherenkov counters with using this silica aerogel could be applied as a new gamma detector of positron emission tomography. Since a Cherenkov photon is emitted in a radiator of n>1.156 for a 511 keV electron and that of n>125 for 341 keV one (compton edge of 511 keV gamma), only photoelectric effect could be detected. The time resolution of a Cherenkov counter is essentially more superior than that of a scintillation counter. The TOF measurement could determine the three dimensional coordinate of the annihilation position of the positron with sufficient resolution. The production method of silica aerogel with such refractive index is shown. The expected properties of this Cherenkov counter are discussed. A simple proof-of-principle experiment confirms the feasibility of the new PET detector.
2
Study of a threshold Cherenkov counter based on silica aerogels with low refractive indices
I. Adachi, T. Sumiyoshi, Kohei Hayashi et al. · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · 1995 · 117 citations · Full text
Natural Sciences, Materials Engineering, Materials Science +9