Concepedia

Publication | Open Access

Evaluation of the pocH-100iV DIFF hematology analyzer for use in horses and cattle

10

Citations

1

References

2009

Year

Abstract

The results of the analysis of equine and bovine blood samples with the automated pocH-100iV DIFF hematology analyzer were compared with the results obtained with reference methods or other analyzers (Vet ABC, Coulter Counter ZF, Coulter LH 750 analyzer). For equine blood and most parameters in bovine blood good to excellent correlations between methods and analyzers were obtained. For bovine blood good to poor correlations and significant differences were obtained between the pocH-100iV DIFF and other methods or analyzers mainly for hematocrit and hemoglobin determinations and platelet counts. Overall the pocH-100iV DIFF seems to be a reliable and user-friendly analyzer.