IEEE Symposium Conference Record Nuclear Science 2004. · 2005 · 11 citations · 15 references
EngineeringMaterials EngineeringMaterials ScienceDiagnostic ImagingRadiation DetectionRadiation TransportDetector TechnologyGamma-ray DetectionGamma-ray DetectorsRadiation DetectorsRadiation Imaging DetectorsSemiconductor DetectorsDetector DesignDetector Energy ResolutionDetectorsElectrode GeometryRadiation Detection EfficiencyRadiation Energy ResolutionScatter RejectionSmall Animal Imaging
Monolithic CdZnTe detectors are promising for medical and small animal imaging because of a good energy resolution that allows both multi-drug diagnostic and scatter event rejection. Optimizations of electrode geometry and detector thickness are used to improve detection efficiency as well as energy resolution for various material transport properties, with the ULYSSE simulator. As an example, for a 5 mm thick detector with (mutau) <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">e</sub> =3.10 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">-3</sup> cmW, a 1.8 mm pitch gives the best simulated energy resolution on the conventional spectrum at 122 keV. Nevertheless, even with an optimized detector, electronic correction methods contribute to reduce tailing. Correction using cathode and anode signals have been tested with HPBM CdZnTe and THM CdTe:Cl detectors. The appropriate method to correct energy seems to depend on material. This paper also presents experimental results on charge sharing and loss